DP Transmitter Calibration: Safe Isolation to 5-Point Test

Do the job like a technician: 1 Safety, 2 Pick tools, 3 Connect, 4 Calibrate (5 points up and down), 5 Restore. Click the valves on the picture, pick the tools, press the buttons. Mistakes show what really happens. Use Next step to move on.

Procedure map (click a step)

Hint

1 Safety first

Tick the three safety items, then click the valves on the picture in the safe order.

Live bench readings

Applied pressure
Standard mA
Multimeter (cross-check)
Difference
HART digital PV
mA converted to %

Results table

PointIdeal mAUp mADown mAErr upErr dn
Error in % of span. Blue = rising, orange dashed = falling, green band = allowed error. A flat line near zero means a good transmitter.

Calibration certificate

As-found worst error
As-left worst error
Up/down difference (hysteresis)
Standard ratio (TUR)
Result
Typical practice (follow your site procedure): traceable standards, TUR 4:1 or better, 5 points rising then falling, record as-found before any adjustment, then as-left after.

Which instrument for which job

InstrumentUse it forRemember
Hand pump + test gaugeQuick field checkNot accurate enough to calibrate
Pressure calibrator + moduleStandard in the fieldPick a module near the transmitter range
Deadweight testerLab referenceVery accurate, slow, needs a stable bench
Pressure controllerAutomatic, repeatable testsCostly, needs clean gas supply
Loop calibrator (mA)Measure and power the loopUse as the mA standard
MultimeterCross-check the mANever the main standard; always in series
HART communicatorTrim, re-range, read digital PVCompare PV with mA to find a DAC error